Form Plug-and-play starting point Equipment Qualification

Form: Instrument Group (A/B/C) Classification Determination Record

A plug-and-play form to document a USP <1058> instrument classification decision: intended use, the measurement and software questions, the resulting Group A/B/C, the qualification scope it drives, and QA approval, with a filled specimen.

Document type: Form

Read and copy the template below into your own quality system. It is a generic starting point for your own internal use, provided as is, with no warranty; see the Terms and License. Adopting it does not by itself create compliance.

This is a ready-to-use form for recording an instrument classification decision under USP General Chapter <1058>. Classification is a documented decision, not an opinion, and this form is the record an auditor asks for when they want to see why a given instrument carries the qualification burden it does. Replace every <<FILL: ...>> placeholder with your specifics and route the completed form through your normal review and approval. A worked filled specimen follows. Verify cited regulations against the current source. This is general guidance to adapt, not regulatory advice.

FieldEntry
Instrument / asset ID<<FILL>>
Instrument type and model<<FILL>>
Location / laboratory<<FILL>>
Intended use (methods and decisions it supports)<<FILL: be specific, e.g. related-substances by gradient HPLC feeding release>>
Classification date<<FILL: YYYY-MM-DD>>
Classified by<<FILL: name, role>>

Classification questions

Answer in order. The first Yes that lands the instrument in a group sets the group, subject to the intended-use confirmation at the end.

#QuestionAnswerNotes
Q1Does the instrument produce a measurement or result that feeds a GxP decision?<<FILL: Yes/No>>If No, it is Group A
Q2Does software acquire, process, or store that result such that you would need it to reproduce or defend the result?<<FILL: Yes/No>>If No (and Q1 Yes), it is Group B
Q3Is software integral to the measurement (acquisition, integration, processing, or storage of the record)?<<FILL: Yes/No>>If Yes, it is Group C
Q4Confirm against intended use: could the same hardware be used a more demanding way (kinetic, curve-fit, networked storage)?<<FILL: Yes/No>>If Yes, classify against the more demanding use in use

Determination

FieldEntry
Assigned group<<FILL: A / B / C>>
Rationale (why this group for this intended use)<<FILL>>
Qualification scope this drives<<FILL: e.g. Group A conformance check; Group B IQ/OQ + calibration; Group C full IQ/OQ/PQ + CSV>>
Part 11 / Annex 11 features to assess (if any)<<FILL: e.g. USB export, data logger, none>>
Related qualification protocol references<<FILL: DQ/IQ/OQ/PQ doc numbers>>

Field definitions

FieldFormatRequiredWhoWhen
Instrument / asset IDtextYesInstrument ownerAt classification
Intended usetextYesInstrument ownerAt classification
Q1 to Q4 answersYes/No + notesYesInstrument ownerAt classification
Assigned groupA/B/CYesInstrument ownerAt classification
RationaletextYesInstrument ownerAt classification
Reviewer / approver (QA)signature, dateYesQAAt approval

Approval

RoleNameSignatureDate
Classified by (instrument owner)<<FILL>>
Reviewed / approved by (QA)<<FILL>>

Reclassify and re-approve whenever intended use changes, because a wider or tighter use can move the group and can invalidate the basis of the existing qualification.


Filled specimen

Illustrative example, values invented.

FieldEntry
Instrument / asset IDLAB-0233
Instrument type and modelUV/Vis spectrophotometer, benchtop, with local PC software
Location / laboratoryQC chemistry
Intended useContent-uniformity absorbance readings at a single wavelength, software acquires and stores the spectrum, result feeds release
Classification date2026-07-03
Classified byA. Rao, QC analyst

Questions: Q1 Yes (absorbance result feeds release). Q2 Yes (software acquires, processes, and stores the spectrum; the stored spectrum is needed to defend the result). Q3 Yes (software integral). Q4 Yes (unit is networked and stores records).

Determination: Group C. Rationale: although the read is a single wavelength, the software acquires, processes, and stores the spectrum and result, and USP <1058> lists UV/Vis among Group C examples; the stored electronic record is part of the measurement system. Qualification scope: full IQ/OQ/PQ plus computer system validation, audit trail verification, individual accounts, raw data retention. Part 11 features to assess: audit trail, access control, electronic records storage. Related protocols: IQ-0233, OQ-0233, PQ-0233.

Approval: Classified by A. Rao 2026-07-03; reviewed and approved by QA (M. Chen) 2026-07-04.

A contrasting note: a stripped-down UV/Vis with only a digital readout and manual transcription of a single number, no software storing the spectrum, could be justified as Group B on this same form, with the rationale recorded. The form is where that justified deviation from the default lives.