Log Plug-and-play starting point Data Integrity

Log: Manual Time-Check for Standalone GxP Instruments

A plug-and-play log for the periodic manual clock check on standalone or legacy GxP instruments that cannot join network time: field definitions, the tolerance and out-of-tolerance action, retention, and a filled specimen row set, with the regulations it satisfies.

Document type: Log

Read and copy the template below into your own quality system. It is a generic starting point for your own internal use, provided as is, with no warranty; see the Terms and License. Adopting it does not by itself create compliance.

This is a ready-to-use log for the periodic manual time check on standalone or legacy GxP instruments that cannot synchronize to the network time service. A standalone instrument whose clock nobody checks is a recurring inspection finding; a tight, evidenced manual check is the compensating control that closes it. Replace every <<FILL: ...>> placeholder, keep one row per check, and maintain the log under document control. A field-definition table, the blank log, instructions, and a filled specimen follow. Verify each cited regulation against the current source. This is general guidance to adapt, not legal or regulatory advice. This log is the record referenced by the SOP: Time Synchronization and System Clock Control for GxP Systems.

Log header

FieldEntry
Log number<<FILL: LOG-ID>>
Instrument / ID<<FILL: make, model, asset ID>>
Location<<FILL: room / lab>>
Check frequency<<FILL: e.g. weekly, or before each use>>
Tolerance<<FILL: e.g. within 60 seconds of the reference>>
Reference used<<FILL: e.g. a synchronized wall display / NTP-synced workstation clock>>

Field definitions

FieldFormatRequiredWhoWhen
Date/time of checkDate + timeYesOperatorAt each scheduled check
Reference time (source)Time + named sourceYesOperatorAt the check
Instrument time displayedTimeYesOperatorAt the check
DifferenceSeconds (signed)YesOperatorAt the check
Within tolerance?Yes / NoYesOperatorAt the check
Action if out of toleranceText + referenceIf NoOperator / QAAt the check
OperatorInitials / signatureYesOperatorAt the check
Reviewed byInitials / signatureYesSupervisor / QAAt review

Blank log

Date/time of checkReference time (source)Instrument timeDifference (s)Within tolerance?Action if outOperatorReviewed
<<FILL>><<FILL>><<FILL>><<FILL>><<FILL: Y/N>><<FILL>><<FILL>><<FILL>>
<<FILL>><<FILL>><<FILL>><<FILL>><<FILL: Y/N>><<FILL>><<FILL>><<FILL>>

Instructions

  1. Perform the check at the defined frequency (and before use if that is your rule).
  2. Read the reference time from the named synchronized source and the instrument’s displayed time at the same moment.
  3. Record the signed difference in seconds (instrument minus reference).
  4. If the difference exceeds the tolerance, do not simply reset the clock and move on. Record the out-of-tolerance value, assess whether any GxP data was generated since the last in-tolerance check that could be affected, correct the clock only under the controlled procedure, and route to QA per the deviation process if data may be affected.
  5. A supervisor or QA reviews the log on a defined cadence and at the point any GxP result from this instrument is reviewed.
  6. Retain the log for not less than <<FILL: retention period>>.

Out-of-tolerance action (summary)

SituationAction
Within toleranceRecord and continue
Out of tolerance, no data since last good checkCorrect under procedure, note cause, continue
Out of tolerance, GxP data generated since last good checkQuarantine the affected data, assess impact, raise a deviation, correct under change control

References

21 CFR Part 11.10(e) (computer-generated, time-stamped audit trails). 21 CFR 211.68 (automatic, mechanical, and electronic equipment), 211.194 (laboratory records). EU GMP Annex 11 (computerised systems). MHRA GxP Data Integrity Guidance and Definitions (2018). PIC/S PI 041, Good Practices for Data Management and Integrity.

Confirm the current version of each reference before issue.


Filled specimen

Illustrative rows for a standalone plate reader checked weekly against an NTP-synced workstation, tolerance 60 seconds.

Date/time of checkReference time (source)Instrument timeDifference (s)Within tolerance?Action if outOperatorReviewed
2026-06-08 08:0508:05:10 (WS-QC-04, NTP-synced)08:05:22+12YesnoneJ.K.R.G. 2026-06-08
2026-06-15 08:1008:10:03 (WS-QC-04, NTP-synced)08:11:47+104NoNo GxP runs since last check; clock corrected under CR-2026-0455; DEV not requiredJ.K.R.G. 2026-06-15

In the second row the drift exceeded tolerance, so the operator did not just fix it silently: they confirmed no GxP data was affected, corrected under a change record, and had it reviewed. Had a run occurred in the drift window, the row would instead reference a deviation and an impact assessment.

Common inspection findings this log prevents

  • A standalone instrument drifting for months with no evidence anyone ever checked its clock.
  • A clock quietly reset with no record and no assessment of data generated during the drift.
  • Manual checks performed but never reviewed, so an out-of-tolerance reading goes unnoticed.

How to adapt this log

  1. Set the instrument, frequency, tolerance, and named reference source.
  2. Decide your rule for “before each use” versus a fixed cadence based on the instrument’s drift and criticality.
  3. Point the out-of-tolerance action to your real deviation and change-control procedures.
  4. Confirm every regulation against the current published version before issue.